Sorvall Lynx 6000 following specs are required for specific laboratory use: swinging bucket and fixed angle rotors that meet our users' needs; can reach higher speeds than comparable models on the market; offers a temperature set range of -20 °C up to +40 °C, with a temperature control accuracy of +/-2 °C; allows setting and controlling in both rpm and RCF directly for any selected rotor; centrifuge door opens fully and automatically, even when a user's hands are full, using an easy to see and reach button near waist height; is designed with a space for feet to be placed under the front of the unit to allow the user to stand closer to the instrument for easier lifting of rotors in and out of the unit; has a "lift over" profile no greater than 860 mm (33.9 in), providing a safe working height for instrument users; runs quietly, at less than 59 dBA; Select rotors must be certified as autoclavable to 121 °C; must meet our space constraints by having a floor-mounted configuration, not exceeding dimensions (H x D xW) of 930 mm x 805 mm x 700 mm (36.6 in x 31.7 in x 27.6 in).