The Contractor shall provide a Steritest Symbio ISL Pump or equivalent. The instrument shall meet or exceed the following requirements:
- The instrument shall consist of a peristaltic pump compatible with Steritest canister devices.
- The pump shall have a table integrated design to be installed into a circular cut-out, measuring 210mm, on a horizontal worksurface of a Getinge Isoflex-S 3-Glove Isolator, serial number #3089.
- The pump head must be capable of automatic opening and closing.
- The pump housing is to be constructed of 316L stainless steel.
- The pump housing and components to be resistant to vaporized hydrogen peroxide gas.
- The peristaltic pump must have an adjustable speed with a maximum speed of at least 200 rpm.
- Pump shall be equipped with a keypad for selecting instrument settings.
- Pump shall be equipped with a timer function.
- Pump shall be equipped with a pressure control system.
- Pump shall be equipped with a color LCD user interface screen.
- Pump shall have the ability to store at least 200 filtration protocols.
- The pump body must not exceed the following measurements: 10 inches in width, 13 inches in depth and 8 inches in height.