Non-Invasive and Hemodynamic Monitoring System:
1. Must provide advanced hemodynamic monitoring parameters.
2. Must have the capability to utilize minimal and noninvasive hemodynamic monitoring in the same platform.
3. Must be able to measure blood pressure, cardiac output, stroke volume, SVV and other related hemodynamic parameters.
4. Must have real time reconstruction of the brachial blood pressure.
5. Must have a heart preference sensor to compensate for potential errors due to difference in height between the finger and heart level.
6. Must be compatible with EV1000 Clinical Platform.
Power Requirements:
1. Electrical Power: 100 to 240 Volts AC 50/60Hz
2. Power Consumption: 100VA, Maximum
3. Protection: Class I Equipment
Clinical Platform:
Must provide accurate hemodynamic monitoring for CI, SVV, SV, SCV02
FLORTRAC Cable 16FT:
Must Connect the patient to the Flortrac Sensor
TABLE STAND:
Must allow for the Flortrac sensor to be positioned , leveled with the patient's phlebostatic axis for accurate measurements