Handpiece must be compatible with Joule Laser System ECN# 086210, Sciton. Handpiece must generate Erbium:YAG laser 2940 nm wavelength. Handpiece software must be compatible with Sciton Joule Laser System computer. Handpiece must be able to generate a laser for ablation to the depths ranging from 25 micrometers to1000 micrometers Handpeice must be able to generate laser in a profractioaned pattern ranging from 1.3 mm x 1.3 mm to 20 mm x 20 mm in size. Handpiece must be able to adjust treatment area percentage from 5.5% to 22% with the generated laser. Handpiece must be capable to genererate laser for coagulation depth of 50 micrometers to 150 micrometers. Vendor must provide Handpiece software install. Vendor will provide initial set up of handpiece to laser system. No grayed/refurbished. New Equipment Only. Operator/User Manual 2 each. Module - (High speed 430micron with randomized pattern software) - requires 2940 module. The scanner handpiece needs to be compatable with the M8527M Joule Laser system. The handpiece allows the use of the ER:YAG 2940 nm laser to penetrate deep into the dermis to treat disfiguring scars. The laser system with handpiece treats disfiguring scars.