FTIR spectrometer specifications:
• Bench top instrument capable of providing mid infrared range spectra (for sample analysis) 7800-350 cm-1.
• 0.4 cm-1 or better resolution
• 0.02 cm-1 wavenumber accuracy
• A interferometer with a non-air bearing (Michelson design)
• A user replaceable laser and source
• Rapid scan capability with low signal to noise
• Auto align capability
• Adaptable to other types of beam splitters, accessories (i.e. an IR microscope) and detectors
• High performance DTSG detector
• System electronics onboard, USB computer interface with full analytical capabilities and operating software
• Software controls for setup and operational parameters, data generation in traditional IR formats, data processing tools, peak ID, peak quantification, other peak analysis tools, spectral search, Data conversions, data correction tools, atmospheric correction, and multi format data conversion tools.
• Compatibility with Nicolet "smart" accessories (i.e. ATR attachments and sample holders)
• Diamond ATR accessory