The 57 RQS at RAF Lakenheath, United Kingdom, is looking for an electrotherapy ultrasound system. The system shall contain four (4) modules including E-Stim, Ultrasound, Laser and EMG. It shall be able to administer combination therapy by using more than one module simultaneously.
In addition, the unit shall possess:
-12 different waveforms
-Meet FDA consensus standards, especially in regards to the output parameters
-User-friendly interface
-An output accuracy of +/- 20%, 10% maximum
-Duty cycles of 10%, 20%, 50% and Continuous
-Head warming feature to increase patient comfort
-Laser component shall have pulse frequency up to 10,000 Hz with an Output Accuracy of +/- 20%
-Diodes that create wavelengths between 670-950nm and have the ability to reach outputs as low as 100mW and as high as 1400 mW or higher
-Diode cooling feature to decrease accidental burning of multiple patients
-EMG component shall be surface (sEMG) and have at least two (2) channels
-Integrated electrical lead management system
-A castered cart for ease of mobility
The unit will need to fit into a 30"x30" space restriction.