This order is modified to add additional specifications to the description.
sterilizer must be recessed with the controls bumped out past the wall for access
Single door that is vertical sliding
chamber size between 18" × 18" × 36" and 21'' x 21'' x 38''
Control Screen with at least 15 programmable cycles
Controls: 110 Volts, 60 Hz with what is already set up through the building
Controls Mounted on Left Side or Right Side
Steam Generator - 25 KW to 30 KW since we have no house steam
Generator: 230-240 Volts, 3 Phase
Automatic Generator Blowdown
Automatic Jacket Blowdown for Liquids Cycle
End-of-Cycle Vacuum Drying
Any piping that is needed to connect the generator/boiler to the sterilizer
A Reverse Osmosis Water Filtration System: between a 25 and 40 gallon tank
Air Compressor
Emergency Stop Button
2 Shelves - 1 Stationary Bottom Shelf & 1 Removable Upper Shelf
1-2 Year Preventative Maintenance Agreement
De-installation and removal of existing sterilizer: Including de-installation, disconnection of all existing
utilities, and removal of sterilizer from customer site.
Installation: Includes receiving sterilizer, uncrating, setting into place, leveling, final
assembly, final utility connections, start-up, and user training.