The USDA, Agriculture Research Services in Albany, California has a requirement for one each Multimode Microplate Reader. Functions: Instrument is a dual-monochromator, multi-detection microplate reader with a 6-384 microplate reader capability. Detection modalities include absorbance(UV-Vis), fluorescence intensity (F), time-resolved fluorescence (TRF) and luminescence (Lum).
Performance: Microplate Formate: 6, 12, 24, 48, 96, 384 well
Temperature Control: 4 deg. C above ambient to 60 deg. C
Wavelength Selection: Monochromator, tunable in 1.0 nm increments.
Endpoint Reading: All modes.
Kinetic Reading: All modes.
Spectral Scanning: All modes.
Well Scanning: Abs, FI, TRF, Lum.
Absorbance photometric performance
Wavelength Range: 200-1000 nm
Typical read time: 18s (96-well), 49s (384-well)
Fluorescence intensity performance
Reading Capabilities: Top or bottom of a microplate.
Wavelength Range: 250-850 nm.
Bandwidth (Ex, Em): 9 nm, 15 nm.
Typical read time: 15s (96-well), 45s (384-well).
Fluorescence polarization performance
Wavelength Range: 400-750 nm.
Bandwidth (Ex, Em): 9nm, 15 nm
Typical read times: 42s (96-well), 120s (384-well)
Time-resolved fluorescence performance
Reading Capabilities: Top or bottom
Wavelength Range: 280-850nm
Bandwidth (Ex, Em): 9nm, 15nm
Precision Data Collection: 1-100 flashes, delay of 0-60 usec before read, integration time selectable between 50-1500 usec.
Sensitivity: 250 Fm europium in 96 or 384 wells with top-read. Typical read time: 17s (96-well), 48s (384-well)
Luminescence performance:
Wavelength Selection: All wavelengths or with selected wavelengths
Wavelength Range: 250-850 nm
Typical read times: 2min (96-well), 7 min (384-well)
Dimensions: 35 cm (H) x 80 cm (W) x 50 (D) or smaller
Electrical: 120 VAC, >8 A, 50/60 Hz
Miscellaneous: Supplied software must provide data acquisition, analysis and management capabilities, allowing cross-plate analysis and custom calculations.
Bid Protests Not Available