The US Geological Survey, Central Energy Resources Science Center (CERSC), Energy Resources Program, Denver, CO has a requirement for a new Inductively Coupled Plasma Mass Spectrometer (ICPMS) capable of major and minor elemental analysis of aqueous and solids materials to replace an existing Inductively Coupled Plasma Mass Spectrometer (ICPMS).
Small business sources are sought for a new Inductively Coupled Plasma Mass Spectrometer (ICPMS) capable of major and minor elemental analysis of aqueous and solids materials
REQUIREMENTS:
1) The ICPMS offered to the USGS must include liquid sample introduction system, RF plasma ion source, quadrupole based reaction cell to eliminate spectroscopic interferences using either collision gases or pure reaction gases such as ammonia and oxygen for reaction cell, quadrupole mass filter, quadrupole ion deflector, simultaneous dual stage discrete dynode detector and a multiple cone interface, all under computer control.
2) Quadrupole must be able to scan a minimum of 5000 amu/sec to achieve maximum productivity from transient signals when jumping between light masses and high masses.
3) Quadrupole resolution settings are dynamically scanned as each element is analyzed.
4) High mass range must be at least 285 amu or greater.
5) All analytical system operations from component optimization to methods development to calibration to analysis to reports must be able to be performed using a single software program.
6) System must be fully compliant with US EPA methodology for Method 200.8 and Method SW 846-6020.
7) ICP-MS system must have a full color plasma view window for useful visuals on the sampler cone, plasma color and injector tip.
8) Close coupled, three-channel integral peristaltic pump, computer controlled, must be integrated within the ICP-MS system.
9) Fully automated X, Y, Z torch alignment.
10) The RF generator which electrically decouples the plasma from the ion optics and allows independent adjustment of the ion optic parameters and the plasma conditions.
11) A multiiple cone design consisting of sample, skimmer and additional skimmer cones with all cones at ground potential.
12) System must incorporate a universal reaction or collision cell offering three modes of operation: standard mode, collision cell mode and reaction cell mode with the ability to utilize a wide variety of gases (including pure reactive gases such as ammonia or oxygen) with no degradation to the gas plumbing, turbo pump or cell.
13) The ion detector must be a simultaneous dual-stage discrete dynode electron multiplier, providing element concentration calibration over 10 orders of magnitude (from 0.1 cps to > 1e9 cps) dynamic range in a single scan using both analog and pulse ion-counting mode, and offer protection against overload in both pulse counting and analog modes.
14) The dual-stage detector assembly must come standard with the system.
15) The system must include a random access auto sampler capable of holding 150 or more 15 mL sample vessels.
16) The system must be capable of being interfaced with ESI Auto samplers and FAST systems.
17) The level of oxide and doubly charged ions generated by the plasma must be less than 3% percent for CeO+/Ce+ and less than 3% for Ba2+/Ba+ using the sample introduction system and cones provided with the instrument
The USGS is in search of small business firms who are able to provide the above mentioned supply/service. Small business firms who feel they can furnish the required supply/service or its functional equivalent (based on criteria specified herein) are invited to submit in writing an affirmative response to this announcement. An affirmative response would include literature, brochures and such other materials, which correspond to the required supply/service stated herein. This information will be used to determine if competitive opportunities exist.