SALIENT CHARACTERISTICS- TOMOTHERAPY SYSTEM
The tomotherapy system requires a detector, a phantom, and compatible software that meets or exceeds the following approximate specifications.
a. DETECTOR
Device type: Vented ionization chambers
Detector design: cubic
Number of detectors: 729
Detector size: 0.5 cm x 0.5 cm x 0.5 cm
(0.125 cmó)
Detector spacing: 10 mm center-to-center,
5 mm edge-to-edge
Max. field size: 27 x 27 cmò
Max. measurement points: 2916 (four measurements)
Reproducibility: ?ñ 0.5%
Dead time: zero
Repetition rate: 200 ms
Measured quantities: absorbed dose to water (Gy),
absorbed dose rate to water (Gy/min)
Resolution: 1 mGy or 1 mGy/min
Measurement range: (0.5 ... 45) Gy/min
Reference point: 7.5 mm below the surface of the array
Housing material: GPR
Dimensions: 30 cm x 42 cm x 2.2 cm
(W x D x H)
Weight: approx. 5.7 kg
Power supply: (90 ... 265) VAC; (50 ... 60) Hz
PC connection: RS232, Ethernet
b. PHANTOM
Design: Octagon-shaped solid body phantom with two exchangeable bottom parts (LINAC phantom with air cavity, CT phantom)
Dimensions: diameter 32 cm, length 32 cm
Weight: approx. 24 kg
Material: Polystyrene (water equivalent within ~ 2%)
Density: 1.04 g/cm3
Extent of supply: LINAC & CT phantom, positioning device, merge 729 insert
c. SOFTWARE COMPATIBILITY: The included tomotherapy system software must be compatible with the following computer system specifications.
Operating system: Microsoftî Windowsî
(XP Professional,
Vistaî Business x32/x64,
Windows 7 Professional x32/x64)
Processor: Multi-core processor
Memory (RAM): 2 GB
Hard disk: 500 MB of free space for application software
600 MB of free space for .NET Framework 4.0+
Screen resolution: 1280 x 1024
Interfaces: Network interface
Other: Windowsî Internet Explorerî 8.0 or higher, Adobeî Readerî 7.0 or higher