Request for Information: Anesthesia Machine and Vaporizers
This Request for Information Notice is for informational and planning purposes only and shall not be construed as a solicitation or as an obligation or commitment by the Government at this time. This notice is intended strictly for market research. The purpose of this Request for Information notice is to determine interest and capability of Other than Small Businesses and Small Businesses, including VIP-Verified Veteran-Owned Small Business (VOSB) or Service-Disabled Veteran-Owned Small Business (SDVOSB) prospective contractors relative to the North American Industry Classification System (NAICS) Code 339112 (Surgical and Medical Instrument Manufacturing).
The Clement J. Zablocki VA Medical Center located in Milwaukee, WI must replace its current anesthesia machines and vaporizers. The requirement is for equipment on an IDIQ/BPA-type contract with a five-year period. Draeger, Inc. anesthesia and vaporizer equipment is known to meet the Government s needs. The Government is requesting information from other vendors with Federal Supply Schedule (FSS) or open market products that can meet the following specifications/capabilities:
Line Item
Item Number
Item Description
Qty Year 1
Qty Total 5-Year Period
0001
MK06000
Anesthesia Machine - Perseus A500
1
28
0002
M36700/OPC5638
D-Vapor 3000 (Desflurane)
1
14
0003
M36500/OPC563
Vapor 3000 (Sevoflurane)
1
13
0004
M36500/OPC5639
Vapor 3000 (Isoflurane)
1
28
0005
6872020
Infinity ID WaterLock 2
1
28
0006
M36049
CASTrGARD, small (set of 4)
1
28
0007
MQ09298
Seminar Perseus A500
1
5
0008
MQ00922
3rd Party Labor
1
5
Anesthesia Machine. Must:
Support adult, pediatric and neonatal patients
Have less than 1 minute fully automatic start-up including system test upon startup
Minimum dimensions: 35 in length X 45 in width
Have battery backup for a minimum of 150 minutes
Alert staff on display if accessories need to be replaced or filled (i.e breathing hoses, vaporizers)
Easily export and import data via USB
Have flexible breathing bag arm
Contain a low or minimal flow anesthesia decision-support tool
Be compatible with Picis anesthesia record system for serial data collection
Be compatible with current vaporizers or replacement of equal vaporizers to be included:
Draeger, Inc. D-Vapor 3000 (Desflurane) model #: M36700 (Qty 14)
Draeger, Inc. Vapor 3000 (Sevoflurane) model #: M36500 (Qty 15)
Be able to have PC, monitor and UPS for Picis anesthesia record keeper mounted to machine
Ventilator
Able to recognize and support patient's breathing efforts accordingly
Allows manual ventilation in case of loss of power
Ventilation modes needed
Manual/Spontaneous ventilation with continuous positive airway pressure (CPAP)
Volume Control CMV (VC-CMVs)
Pressure Control CMV (PC-CMVs)
Pressure Control SIMV+ (PC-IMVs,s)
Volume Control CMV/AF (PC-CMVa)
Volume Control SIMV/AF (PC-IMVa,s)
Volume Control SIMV/AF/PS (PC-IMVa,s)
Mechanical gas mixer
Gas flow measured with electronic flow sensors
Controls must be able to work in absence of electrical power
Gas analyzer
Must analyze inspired and expired concentration of O2, CO2 and anesthetic gases
Must return the sample to the breathing system
Manual pressure gauge
Electronically displayed
Continues to function in the absence of power
Touchscreen colored display 15 15.5
Display is customizable to user needs
Displays at least three waveforms, parameters and mini trends simultaneously
Displays status of gas (monitors gas consumption), airway pressure and power supplies
Illumination of workstation is controlled on main display
Workstation
is well lit to perform clinical tasks
Work surface must be at least 14 (L) X 36 (W)
Writing surface that is at least 11 in length X 21 in width
includes storage large enough to store a laptop, at least 2 (H) X 11 (L) X 14 (W)
Includes integrated power outlets
Main switch for all outlets
Circuit breakers for each outlet
Vaporizers
D-Vapor Desflurane Vaporizer
Vapor 3000 Sevoflurane Vaporizer
Vapor 3000 Isoflurane Vaporizer
Maximum liquid filling capacity: 300mL
Internal battery provides 3 to 5 minutes back-up power (D-Vapor Desflurane Vaporizer only)
Illuminated control dial and filling level indicators for visibility in dark environments
No calibration required
Automatically identified by anesthesia machine
Type and control dial setting of each vaporizer needs to be visible on screen
Accessories
Infinity ID Waterlock 2
CASTrGARD
Training and Installation
In-service training for two (2) Biomed staff and anesthesiologists for at least two weeks, Monday-Friday 6:30am to 3pm
Installation of equipment provided
Note that this requirement is determined to be ineligible for set-aside; however, interested Other Than Small Businesses and Small Businesses including VIP-Verified VOSBs and SDVOSBs are required to submit an e-mail delineating:
Qualifications, capabilities, and experience for providing this product;
Business size/Socioeconomic status;
If the product is on the Federal Supply Schedule or the Open Market; and
Memo or correspondence which authorizes distributor to distribute the manufacturer s products:
Gray market items are Original Equipment Manufacturer s (OEM) goods sold through unauthorized channels in direct competition with authorized distributors. This procurement is for new OEM medical supplies, medical equipment and/or services contracts for maintenance of medical equipment (i.e. replacement parts) for VA Medical Centers. No remanufactures or gray market items will be acceptable.
Vendor shall be an OEM, authorized dealer, authorized distributor or authorized reseller for the proposed medical supplies, medical equipment and/or services contracts for maintenance of medical equipment (i.e. replacement parts), verified by an authorization letter or other documents from the OEM, such that the OEM s warranty and service are provided and maintained by the OEM. All software licensing, warranty and service associated with the medical supplies, medical equipment and/or services contracts for maintenance of medical equipment shall be in accordance with the OEM terms and conditions.
The delivery of gray market items to the VA in the fulfillment of an order/award constitutes a breach of contract. Accordingly, the VA reserves the right enforce any of its contractual remedies. This includes termination of the contract or, solely at the VA s election, allowing the Vendor to replace, at no cost to the Government, any remanufactured or gray market item(s) delivered to a VA medical facility upon discovery of such items.
This is not a Request for Proposal or an announcement of a solicitation and no solicitation package exists at this time. Your responses will be used solely for market research purposes of the Government. The Government will not pay for any materials provided in response to this notice and submissions will not be returned to the sender. VOSBs and SDVOSBs must be registered and verified in VIP for set-aside consideration.
The applicable North American Industry Classification System (NAICS) Code for this requirement is 339112 and the small business size standard is 1,000 employees. Responses due are to be submitted by e-mail to Jim Hedman at
[email protected]. The information requested must be received no later than 09:00 a.m. Central Standard Time on March 8th, 2018.
DISCLAIMER
This RFI is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. In accordance with FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
Bid Protests Not Available