The equipment should provide widest range of treatment options with verified CAD/CAM lab workflows: Crowns, Bridges, Veneers, inlays, onlays, Removable partial dentures (exclusive to TRIOS), Temporary crowns and virtual diagnostic wax-ups (exclusive to TRIOS), Post and core (exclusive to TRIOS), Abutments, implant bridges and bars, and full study model scanning including vestibular and palatal areas
- The equipment should include software necessary for capture of images, editing and laboratory communication with ability of future updates and upgrades
- The system should provide ability to integrate additional componentry including software expansion for diagnostic planning, implant treatment planning and milling services
- The equipment should provide fast, accurate, powder free scanning with option for full arch scanning and articulation
- The equipment should provide option for shade matching
- The equipment should provide ability of scanning all surfaces in oral cavity, including semi-transparent ortho brackets and high gloss metals.
- The equipment should provide removable scanner tips, with ability to sterilize by autoclave
- The scanning tips should provide ability to scan in multiple positions: scanning upwards and downwards.
- The scanner should provide minimum LED light source
- The system should support open platform STL files, which will not limit ability to communicate with other manufacturers milling machines.
- The system should be portable enough for easy transport between multiple dental operatories
- The system should provide equivalent of Pod: Power cable and power supply unit, one Wi-Fi dongle, one charger for TRIOS 3 Wireless batteries including three batteries, USB dongle with software, five TRIOS 3 Scanner Tips including autoclavable mirrors, TRIOS 3 Protection Tip, TRIOS 3 Calibration Tip, TRIOS 3 Color Calibration Kit
- The capturing Pen grip dimension should not exceed: 373 g, 28.22 x 40.03 x 4.87 cm
- Battery minimum requirements: Pen grip: 5-minute
heating, 20-minute continuous scanning
Minimum computer requirements with capacity and power to process high quality images captured by the intraoral scanner:
Display: 15,6" FHD (1920 x 1080) (or more)
Processor: Intelî Core i7-6700HQ or i7-6820HK or i7-7700HQ
Memory: 16GB (or more) DDR4, 2400 MHz (or more)
Disk: 1TB SSD boot disk (or more) (additional spinning disk optional)
Graphics card: NVIDIAî GeForceî GTX 1060M (6GB) or GTX 1070 (8GB) or GTX 1080
Keyboard: Regional
Operating system: Windows 8.1 Pro or Windows 10 Pro (1)
Dim.: 389mm (15.3 ) x 405mm
(12.0") x 25mm (1.0 )
Weight: 3.5 Kg (7.7 lbs.)
PLEASE INCLUDE QUOTE VIA EMAIL ONLY